IT8600 is ITECH latest series of AC/DC electronic loads with power range 0~14.4 kVA and adjustable frequency 45 Hz ~ 450 Hz. The unique oscilloscope waveform display function of IT8600's can display input voltage & current as waveform.

It is equipped with measurement modes for different parameters such as inrush current, peak value, effective value, PF (power factor). Voltage harmonics measurement capacity is up to 50th. The built-in GPIB, LAN and USB communication interfaces are for reliable and fast control.

IT8600 is the perfect solution for testing UPS, inverters, AC power supplies and relevant AC electronic components etc.

Features

  • Frequency range: 45 Hz~450 Hz
  • Power range: 0~14.4 kVA
  • Voltage range: 50~420 Vrms, 15~260 Vrms
  • Current range: 0~1 60 Arms
  • Parallel connection/ three-phase control  *1
  • 7’’ LCD screen
  • Oscilloscope function supporting display of voltage and current waveform
  • High-speed AD sampling, real-time capture waveform
  • Measure Vrms, Vpk, Vdc, Irms, Ipk, Idc, W, VA, VAR,CF, PF and FREQ
  • Measures THD ( V ) up to 50th Harmonic
  • AC electronic load: CC/CR/CP mode
  • DC electronic load: CC/CR/CP/CV mode *2
  • External 0~10 V analog control input, voltage and current analog monitoring function *3
  • OTP, OCP, OVP, UVP and OPP protection function
  • GPIB, LAN and USB communication interfaces and external USB flash disk interface

*1 IT8616 is without this function, IT8617 supports single/three-phase switch output   

*2 Only IT8615 and IT8615L are with CV function   

*3 Only IT8615 and IT8615L have the function of external analog 


 

Applications



Functions and Advantages


Display Multiple Parameters Simultaneously

IT8600 provides 7 inch LCD display screen, easy user interface. Give full consideration to engineers' requirements in different tests, IT8600 not only can display multiple parameters simultaneously,but also has diversified display modes for choice, such as waveform, histogram and list etc.


Harmonic Measuring And Analysis Function

IT8600 provides powerful data measurement function, which can not only support measurement of conventional parameters such as Vrms, Vpk, Vdc, Irms, Ipk, Idc, W, VA, VAR, CF, PF and Freq, but also provide unique voltage harmonic analysis function to verify the harmonic interference of the object (UPS, generators, etc.) to be measured over the grid. The harmonic measurement function supports analysis up to the 50th voltage harmonic and it can display the percentage of each harmonic analysis results in different forms.


Oscilloscope Function

The most unique highlight of IT8600 lies in the oscilloscope display function, which can display the input voltage and current waveform of the device under test
measured. Through the screenshot function key to save the current screen picture to peripheral storage disk by the front USB interface, easy for the second analysis.


Parallel/3-Phase Control

IT8600 provides parallel and 3-phase functions for three-phase and high-power applications. In 3-phase applications, users can make Y or △ connection according to their specific requirements. IT8600 is available for AC 380V input to meet diverse test requirements. ( *Only IT8615 and IT8615L are with the function)



Harmonic measuring and analysis function
IT8615 provides powerful data measurement function, which cannot only support measurement of conventional parameters such as Vrms, Vpk, Vdc, Irms, Ipk, Idc, W, VA, VAR, CF, PF and Freq, but also provides a unique voltage harmonic analysis function to verify the harmonic interference of the object (uninterruptible power supply, generators, etc.) to be measured over the grid. It is equipped with function to conduct analyses for voltage harmonic of as high as 50 orders.

 


Oscilloscope function                                                                                            


The most unique highlight of IT8615 lies in the oscilloscope display function, which can display the input voltage and current waveform of the object to be measured. Under the harmonic measurement mode, the analysis result of the percentage of different harmonics can be displayed in the bar diagram. The diversified display mode can bring a new use experience for the user.                                                   

 


Parallel / three-phase control                                                                                     


IT8615 offered multi-sets parallel,three-phase and parallel three-phase function,widely used in three-phase AC power supplies or high power testing occasions.In parallel three-phase applications,diversify collocations are available for user's actual demand.

 